pH AND mV SIMULATOR HD9609, Merk : DeltaOhm
Keterangan
GENERAL CHARACTERISTICS
The simulator HD 9609 is a portable instrument for checking and calibrating pH and mV measuring instruments. The characteristics of this instrument satisfy any checking and calibrating requirements for both portable and panel-mounted instruments; it may be used in laboratories, in industry or for checks in the fi eld.Despite its many functions, the instrument is simple to use: a large display, with dual indication, and a series of symbols allow it to be used even by unskilled personnel.
The HD9609 sends to output in channel A the simulation of signals of an electrode for measuring pH, ORP, ISFET, in the range:
• 0 to 14 pH, with resolution 0.10 pH;
• ±1999 mV, with resolution 1 mV.
The user may choose between two output impedance values:
• 100 K©, low impedance;
• 1 G©, high impedance.
The simulation of the electrode compensation temperature is manually programmed in the range from -20°C to +150°C, while the temperature is measured in degrees Celsius or Fahrenheit.The pH simulation values may be manually set as desired, in steps of 0.1 or 1 pH. The mV simulation values may be manually set as desired, in steps of 1 or 10 mV.The HD9609 is fed with an ordinary 9Vdc alkaline battery.The electronics are housed in a sturdy ABS case with ergonomic lines.In designing and making the instrument, each detail has been assessed and selected in order to provide an instrument with high performance and excellent long-term measurement stability.On request, the instrument may be certifi ed by a SIT centre or byanother centre recognized by the WECC.
TECHNICAL CHARACTERISTICS
pH simulation: 0÷14 pH
pH resolution: 0.1 pH
pH accuracy 20÷25°C: 0.002 pH
Thermal drift: ±0.0005
pH/°C from -5°C to 20°C and from 25°C to 50°
mV simulation: ±1999 mV
mV resolution: 1 mV
mV accuracy: ±100 µV
Thermal drift mV scale: -199.9 ... +199.9: ±0.01 mV/°C from -5 to 20°C and from 25 to 50°CmV
thermal drift: -1999 ... +1999: ±0.05 mV/°C from -5 to 20°C and from 25 to 50°C
Noise 0÷10 Hz: 1µV peak/peak
Simulation of compensation
temperature: -20 to 150°C (-4 to 302°F)
Output impedance: 100 K© 1%, 1G© 5% (no leading load capacity)
Display: LCD 2 lines, 3 1D2 digits. Figure height approx. 12.5 mm.
Symbols: pH, mV, °C, °F, HI imp., LO imp., 0.1 pH, 1 pH, 1 mV, 10 mV
Signals: LOU, ER1, CALWork temperature: -5 to 50°C (23 to 122°F)Power supply: 9 Vdc alkaline battery. Low battery indication.
Consumption (instrument only): 5 mA lit, 20 µA off
Autonomy: about 200 hours
Dimensions: 187 x 72 x 38 mm.
Weight: 300 gr
ORDER CODES
HD 9609 K: Kit composed of the instrument HD 9609, adapter cables CP 9509BNC, CP 9509 T, carrying case
CP 9509BNC: Adapter cable L = 1 mt, male BNC connector on both ends
CP 9509 T: Adapter cable L = 1 mt, BNC connector on only one end
CP9509S7: Adapter cable L = 1 mt, BNC wall connector one end, S7 male connector on the other end.