Selama 13 tahun BGI PQ100 yang di produksi oleh omniBGI telah terbukti dan teruji untuk melakukan sampling partikel udara di ambient ( PM 10/ PM2,5 / TSP ), dan telah banyak di gunakan di seluruh dunia.

Application

⢠Approved by EPA for PM10 sampling (Designation No. RFPS-1298-124)
⢠Suitable for Lead, Asbestos, Oil Mists, Indoor Air Pollutants and Allergens
⢠Offers well Regulated Air Flow (Mass Flow Control) in the 2 to 25 LPM range

Features

⢠Accurate EPA Specified Flow Rate to ±0.5%
⢠Precision Flow to ±2%
⢠Download Software Included
⢠Complete System is Portable and easily usable Indoors or Out
⢠Weatherproof Enclosure Included
⢠24 Hour Sample on Internal Battery

Universal PQ/TSP Sampling Inlet

The publication of an extensive multi year European aerosol study* of inlet performance confirms the utility of the US EPA louvered inlet. This is essentially the top half of the McFarland Texas A&M Federal Reference Method (FRM) inlet without the impactor and water excluder sections. The Inlet (P/N PQTSP) should be utilized with a minimum of a 19.5â long inlet sample tube (P/N L20) to avoid bluff body effects.

Rack/Support Assembly

This assembly is designed to be rigid and easier to use since the instrument panel is supported at waist height. The legs use the same "snap and lock" assembly developed for the PQ200 (and are, in fact, interchangeable). A large water trap has been included in the new specification for deployment of the instrument in extreme climates.

Solar Panel

The BGI Solar Panel accessory for the PQ200 will be extend its operational time range from 24 hours to an indefinite period depending upon the amount of ambient sunlight available in the region of use. As a practical matter the panel with the ballast battery will provide a minimum of 10-14 days of operation with only minimum ambient light. When interval sampling on three or six day cycles is contemplated continuous operations far from a source of power are achievable

Specifications

PQ100 Sampling Pump
Dimensions : 10 x 6 x 9.7 in (25.4 x 15.2 x 24.6 cm)
Weight : 19 lbs (8.63 kg)

BGI 167 R
Height : 79 in (2 m)
Weight : 34 lb (15 kg)

BGI 16.7 PM10 Inlet Kit: Inlet, Tripod, and Filter Holder
Height : 16 in (40.6 cm)
Diameter : 11 in (28 cm)

Tripod
Height : 68 in (1.73 m)
Diameter : 6 in (15.2 cm) Closed
Height : 39 in (1 m) Disassembled