Jual Produk TSI-Acubalance-8375. Hub : R. Sinaga, Email : pro.teknik@ yahoo.co.id, pro-teknik@ live.com, Telp/ Fax : 021 470 4719, Hp: 0815 1311 6206; 021 93800 487; 08 2124 167 067
Keterangan
TSI-Acubalance-8375.
Specifications
ACCUBALANCE Model 8375
Volume
Range 25 to 2, 500 ft3/ min ( 42 to 4250 m3/ h) capture hood
Accuracy ± 3% of reading ± 7 ft3/ min ( ± 12 m3/ h) at flows > 50 ft3/ min ( > 85 m3/ h)
Units ft3/ min, m3/ h, m3/ min, l/ s
Resolution 1 ft3/ min ( 1 m3/ h)
Velocity
Range ( Pitot Probes) 25 to 8, 000 ft/ min ( 0.125 to 40 m/ s)
Range ( Airflow Probe) 25 to 5, 000 ft/ min ( 0.125 to 25 m/ s)
Velocity Matrix 25 to 2, 500 ft/ min ( 0.125 to 12.5 m/ s)
Accuracy ± 3% of reading ± 7 ft/ min ( ± 0.04 m/ s) at
velocities > 50 ft/ min ( 0.25 m/ s)
Units ft/ min, m/ s
Resolution 1 ft/ min ( 0.01 m/ s)
Pressure
Differential Pressure
± 15 in. H2O ( ± 3735 Pa ) ; 150 in. H2O
( 37.5 kPa) maximum safe operating pressure
Absolute Pressure 15 to 40 in. Hg ( 356 to 1016 mm Hg)
Accuracy ± 2% of reading ± 0.001 in. H2O ( ± 0.25 Pa ) static and differential; ± 2% of reading absolute
Units in. H2O, in. Hg, Pa, hPa, kPa, mm Hg, cm Hg, mm H2O, cm H2O
Resolution 0.00001 in. H2O ( 0.001 Pa) static and differential; 0.01 in. Hg ( 1 mm Hg) absolute
RH
Range 0 to 95% RH temperature/ RH probe
Accuracy ± 3% RH
Resolution 0.1% RH
Temperature
Sensor in Base 40 to 140° F ( 4.4 to 60° C)
Temperature Probe -40 to 250° F ( -40 to 121° C)
Temperature/ RH Probe 14 to 140° F ( -10 to 60° C)
Accuracy ± 0.5 ° F ( ± 0.3° C) from 32 to 160° F ( 0 to 71° C)
Units ° F, ° C
Resolution 0.1° F ( 0.1° C)
Masukan ke Produk RekananKenalkan ke teman Anda