The Setaflash Series 8 ' ActiveCool' is an automated closed cup flash point tester with a temperature range of 10 to 130° C ( 50 to 266° F) with no external cooling. When connected to a suitable chilled water supply, the range is extended down to -20° C ( -4° F) .

There are two modes of operation; Flash / No Flash and Ramp. The ramp mode increases the sample temperature at a rate of 2° C/ min automatically dips the ignitor every 1° C up to 100° C ( every 2° C above 100° C) , until either a flash is detected or the end of test temperature is reached.

* 10° C to 130° C ( -20° C with external water supply)
* Small sample size, 2 or 4ml
* Flash/ No Flash & Ramp modes
* Electric ignitor ( gas available using optional accessory)
* Automatic dipping and flash detection
* ActiveCool electronic Peltier cooling
* 64 Test memory & RS232 interface
* ° C or ° F temperature display

Temperature Range: Air: 10 to 130° C ( 50 to 266° F) Water Assisted: -20 to 130° C ( -4 to 266° F) using 99300-2
Sample Size: 2 or 4ml according to method
Ignitor: Electrical Hot Wire ( optional Gas Flame)
Test duration, Rapid Equilibrium Mode: 1 minute below 100° C, 2 minutes above 100° C, FAME 1 minute or user defined 1 to 99 minutes
Test duration, Ramp Mode: Typically 7 minutes

Methods

ASTM D3278, ASTM D3828, ASTM D7236, ASTM E502, BS 3900 Part A13, BS 6664 Parts 3 & 4, CHIPS Regulations, EPA 1020 A & B, IP 523, IP 524, IP 534, ISO 3679, ISO 3680, UN Class 3 Non-viscous Flammable Liquids

DETAILS OF METHODSCHIPS Regulations Chemicals Hazard Information and Packaging for Supply
EPA 1020 A & B Setaflash Closed-Cup Method for determining ignitability
ASTM D3828 Standard Test Methods for Flash Point by Small Scale Closed Tester
IP 523 Determination of flash point. Rapid equilibrium closed cup method
IP 524 Methods of test for paints. Tests on liquid paints ( excluding chemical tests) . Test for flash/ no flash ( rapid equilibrium method)
ASTM D7236 Flash Point by Small Scale Closed Cup Ramp Method
IP 534 Flash Point by Small Scale Closed Cup Ramp Method
ASTM D3278 Standard Test Methods for Flash Point of Liquids by Small Scale Closed-Cup Apparatus - The test methods are applicable to paints, enamels, lacquers, varnishes, and related products having a flash point between 0 and 110° C ( 32 and 230° F) and Viscosity lower than 150 St at 25° C ( 77° F) .
ISO 3680 Methods of test for paints. Tests on liquid paints ( excluding chemical tests) . Test for flash/ no flash ( rapid equilibrium method)
UN Class 3 Non-viscous Flammable Liquids Small scale flash point testers are universally specified in many international regulations to define and classify " flammable" and " combustible" substances.
ASTM E502 Standard Test Method for Selection and Use of ASTM Standards for the Determination of Flash Point of Chemicals by Closed Cup Methods
ISO 3679 Determination of flash point. Rapid equilibrium closed cup method
BS 6664 Parts 3 & 4 Methods of test for paints. Tests on liquid paints ( excluding chemical tests) . Test for flash/ no flash ( closed cup equilibrium method)
BS 3900 Part A13 Test for flash/ no flash ( rapid equilibrium method)

Size ( HxWxD) 30 x 34 x 38cm
Weight 8 kg
CCCN Code: Tariff 90268020